Télécharger Imprimer la page

Hillrom Allen Advance Table Mode D'emploi page 39

Publicité

INSTRUCTIONS FOR USE
3.2.8
Adjust the Height of the Allen Advance Table
WARNING:
Before you adjust the Allen Advance Table or use the tilt feature, make sure that
the brakes are locked, the patient is secured to the support top and is fully
prepared for the movement, and the Allen Advance Table is free of objects that
could interfere with the adjustment. Otherwise, patient fall, injury, or equipment
damage could occur.
1.
Make sure that—
• The brakes are locked (the locked indicator on the
Allen Advance Table display is green).
• The patient is secured to the support top and is fully
prepared for the adjustment.
NOTE: We recommend that you use the two large security straps to secure the patient to
the support top.
• The Allen Advance Table is free of objects that could interfere with the
adjustment.
WARNING:
Observe lines closely during articulations. Always use good line management
techniques. Failure to do so could cause patient fall, injury, or equipment
damage.
2.
Press and hold the applicable raise or lower control until
the Allen Advance Table is at the correct height.
NOTE: Near the bottom, on the outside of the columns, there
are numbers to help identify the height level.
Document Number: 80028190
Version: E
Page 39
English
Issue Date: 16 MAY 2022
Ref Blank Template: 80025117 Ver. F

Publicité

loading

Ce manuel est également adapté pour:

A-71100-sch