INSTRUCTIONS FOR USE
3.7.2
Fuses
Removal
1.
Make sure the brakes are locked (the locked indicator on
the Allen Advance Table display is green).
WARNING:
Failure to remove power from the Allen Advance Table could cause patient fall,
injury, or equipment damage.
2.
Turn the Allen Advance Table off, and unplug the power
cord from the facility's power source and from the Allen
Advance Table.
3.
Pull the fuse holder from the plug housing.
NOTE: You may need to use a small screwdriver or a pen to pry
the holder from the housing.
4.
Remove the fuses from the holder.
Replacement
1.
Install the new fuses in the holder.
2.
Fully insert the holder into the plug housing.
3.
Plug the power cord into the Allen Advance Table and into an applicable power
outlet, and turn the Allen Advance Table on.
4.
Make sure all Allen Advance Table functions operate correctly.
Document Number: 80028190
Version: E
Page 60
English
Issue Date: 16 MAY 2022
Ref Blank Template: 80025117 Ver. F