INSTRUCTIONS FOR USE
3.2.9
Put the Allen Advance Table in a Trendelenburg Position
WARNING:
Before you adjust the Allen Advance Table or use the tilt feature, make sure that
the brakes are locked, the patient is secured to the support top and is fully
prepared for the adjustment, and the Allen Advance Table is free of objects that
could interfere with the adjustment. Otherwise, patient fall, injury, or equipment
damage could occur.
1.
Make sure that—
• The brakes are locked (the locked indicator on the
Allen Advance Table display is green).
• The patient is secured to the support top and is fully
prepared for the adjustment.
NOTE: We recommend that you use the two large security straps to secure the patient to
the support top.
• The Allen Advance Table is free of objects that could interfere with the
adjustment.
WARNING:
Observe lines closely during articulations. Always use good line management
techniques. Failure to do so could cause patient fall, injury, or equipment
damage.
2.
Press and hold the applicable Trendelenburg or Reverse
Trendelenburg control until the support top is at the
correct angle.
3.
To return to the horizontal position, press and hold the opposite Trendelenburg control
until the lowered column is at the same height as the raised column.
NOTE: Near the bottom, on the outside of the columns, there
are numbers to help identify the height level.
Document Number: 80028190