Télécharger Imprimer la page

B.Braun Aesculap Acculan 4 Mode D'emploi page 7

Masquer les pouces Voir aussi pour Aesculap Acculan 4:

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 30
6.
Working with the device
WARNING
Risk of infection and contamination!
Product is delivered unsterilized!
Sterilize the product before use pursuant to the operating instruc-
tions.
WARNING
Risk of injury and material damage due to accidental activation of the
product!
Products which are not being actively used must be secured against
accidental activation (position OFF).
WARNING
Risk of injury and material damage due to inappropriate use of tools!
Always follow the safety advice and information given in the
instructions for use.
When coupling/uncoupling, handle tools with cutting edges with
care.
WARNING
Damage to the product if dropped!
Use the products only if they are in perfect technical condition, see
Function check
WARNING
Risk of burns to skin and tissue through blunt tools or if product has
not been maintained properly!
Use tools only if they are in perfect condition.
Replace blunt tools.
Maintain the product properly, see maintenance guide.
6.1
System set-up
6.1.1
Connecting the accessories
Combinations of accessories that are not mentioned in the present
instructions for use may only be employed if they are specifically intended
for the respective application, and if they do not compromise the perfor-
mance and safety characteristics of the products.
All
configurations
must
IEC/DIN EN 60601-1. The person connecting the devices with each other
is responsible for the configuration and must ensure that the fundamental
standard IEC/DIN EN 60601-1 or relevant national standards.
Follow the instructions for use of individual accessories.
In case of questions, contact your B. Braun/Aesculap partner or
Aesculap Technical Service, address see Technical Service.
6.1.2
Inserting the rechargeable battery
CAUTION
The product does not function when using the wrong battery!
Do not use battery GA666 (with red base) or GA346.
Product 1 with the battery shaft 4 must be turned upwards and the
sterile funnel 6 (sterile) attached, see Fig. A.
Battery 7 (non-sterile) is inserted into the battery shaft 4 by a second
(non-sterile) person, see Fig. A.
Note
After the battery has been inserted, multiple signals will sound, which indi-
cate that the product is ready to use.
fulfill
the
fundamental
After insertion of the battery, have the sterile funnel 6 (non-sterile)
removed by a second person.
Attach the lid 8 (sterile) in such a manner that it clicks into place with
both release catches 9.
Note
The sterility of the product is only guaranteed with a correctly placed lid.
6.1.3
Intraoperative battery change
WARNING
Risk of injury from hot battery!
Battery may be hot after use in the machine.
Remove the battery with the battery removal device and let it cool.
The battery removal device is used for replacing the battery during surgery
while ensuring that sterile conditions are maintained.
Turn the product 1 with the battery shaft 4 upwards.
Press both safety catches 9 on the lid 8 at the same time and remove
the lid 8.
Attach the sterile battery removal device 10, see Fig. B.
Gently shake the product 1 with the attached battery removal
device 10 with battery compartment 4 facing downwards.
The battery 7 slides easily into the battery removal device 10.
Hand battery removal device 10 including dead battery 7 to a non-
sterile person.
Insert a charged battery 7, see Inserting the rechargeable battery.
6.1.4
Removing the rechargeable battery
WARNING
Risk of injury from hot battery!
Battery may be hot after use in the machine.
Let the battery cool in the machine and then remove.
– or –
Remove the battery with the battery removal device and let it cool.
CAUTION
Damage to the battery caused by knocking it against hard objects!
Only remove battery by tapping the product on the palm of the
hand.
standard
CAUTION
Damage to, or destruction of the batteries due to processing!
Do not sterilize batteries.
Rechargeable battery must be removed after each surgical procedure and
prior to processing of the device.
Note
The battery removal device can be used to make battery removal easier, see
Fig. B.
Turn the product 1 with the battery shaft 4 upwards.
Press both safety catches 9 on the lid 8 at the same time and remove
the lid 8.
Tightly grasp the battery shaft 4 at the lower end.
Tap the battery shaft 4 on the palm of the hand until the battery 7
slides out of the battery shaft 4 and can be removed.
en
5

Publicité

loading

Ce manuel est également adapté pour:

Ga330