Télécharger Imprimer la page

B.Braun Aesculap Acculan 4 Mode D'emploi page 10

Masquer les pouces Voir aussi pour Aesculap Acculan 4:

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 30
en
7.7
Manual cleaning with wipe disinfection
Phase
Step
I
Pre-cleaning
II
Cleaning with enzyme solution
III
Intermediate rinse
IV
Drying
V
Wipe disinfection
VI
Final rinse
VII
Drying
D–W:
Drinking water
FD–W:
Fully de-ionized water (de-mineralized, microbiological, with minimum quality of drinking water)
RT:
Room temperature
*
Suitable enzyme solution: Helizyme, Cidezyme (the latter for validation purposes)
Do not clean the product in a ultrasonic bath and do not immerse the
product in any fluids. Let any fluid incursions drain out immediately,
otherwise there is a danger of corrosion and loss of function.
Phase I
Mobilize non-rigid components during cleaning.
Clean the product under running water, using a suitable cleaning brush
until all visible residues have been removed from the surfaces.
Brush difficult to access surfaces with a suitable plastic cleaning brush
for at least 1 min.
Note
For details on difficult to access surfaces, see Acculan pre-cleaning and
care information TA016000 (available in Aesculap Extranet at
https://extranet.bbraun.com).
Phase II
Follow the operating instructions of the enzyme cleaner with regard to
correct concentration, dilution, temperature and water quality.
Spray products with a pH neutral enzyme solution, let soak in for at
least 2 minutes and then wipe off.
Contamination should be removed with a lint-free cloth or soft brush
moistened with enzyme cleaner.
Flexible components should be rinsed for 20 seconds with the water
pistol (cold water, at least 2.5 bar).
After manual cleaning, check visible surfaces and areas of flexible com-
ponents for residues.
Phase III
Mobilize non-rigid components during cleaning.
Rinse product under running tap water for at least 5 minutes.
If necessary, repeat the cleaning process (phase 1 to 3).
Phase IV
Dry the product in the drying phase with suitable equipment (such as
lint-free cloths, pressurized air).
8
T
t
Conc.
[°C/°F]
[min]
[%]
≥2
RT (cold)
≥2
RT (cold)
0.8
≥5
RT
RT
>1
RT (cold)
0.5
RT
Water qual-
Chemicals
ity
D–W
Until visually clean
D–W
pH-neutral*
D–W
Meliseptol HBV wipes 50 % Propan-1-ol
FD–W
Phase V
Wipe all surfaces of the product with a single-use disinfecting wipe.
Phase VI
Rinse disinfected surfaces after the prescribed reaction time for at least
1 minute under running demineralized water.
Drain any remaining water fully.
Stage VII
Dry the product in the drying phase with suitable equipment (such as
lint-free cloths, pressurized air).

Publicité

loading

Ce manuel est également adapté pour:

Ga330