7.8
Automatic cleaning/disinfection with manual pre-cleaning
Note
The cleaning and disinfection device must fundamentally have a tested
efficacy (such as FDA approval or CE label pursuant to DIN EN ISO 15883).
7.8.1
Manual pre-cleaning with a brush
Phase
Step
I
Rinsing
II
Brushes
D–W:
Drinking water
RT:
Room temperature
Do not clean the product in a ultrasonic bath and do not immerse the
product in any fluids. Let any fluid incursions drain out immediately,
otherwise there is a danger of corrosion and loss of function.
Phase I
Mobilize non-rigid components during cleaning.
Thoroughly clean the product under running water.
Phase II
Mobilize non-rigid components during cleaning.
Brush difficult to access surfaces with a suitable plastic cleaning brush
for at least 1 min.
7.8.2
Mechanical alkaline cleaning and thermal disinfection
Machine type: single-chamber cleaning/disinfection device without ultrasound
Phase
Step
I
Pre-rinse
II
Cleaning
III
Intermediate rinse
IV
Thermal disinfection
V
Drying
DW:
Drinking water
FD–W:
Fully desalinated water (demineralized, low microbiological contamination: drinking water quality at least)
*Recommended: BBraun Helimatic Cleaner alkaline
Insert the product in its correct position in the ECCOS holder.
Screw the rinsing adapter (GA672211) on the product in the mount 14.
Then connect the interior rinse device and connect to the rinse connec-
tor of the cleaning/disinfection device/rinse cart.
Attach the Kirschner wire protection sleeve 15 to a rinse hose.
After automatic cleaning/disinfection, check visible surfaces for resi-
dues and repeat the cleaning/disinfection process as needed.
T
t
Conc.
[°C/°F]
[min]
[%]
RT (cold)
RT (cold)
T
t
[°C/°F]
[min]
<25/77
3
55/131
10
>10/50
1
90/194
5
max.
min. 10min
120/248
Note
The cleaning and disinfection machine used for processing must be serviced
and checked at regular intervals.
Water
Chemistry/Note
quality
until visually clean
D–W
D–W
until visually clean
After manual preparation, check visible surfaces for residue and repeat
the pre-cleaning process as needed.
Note
For details on difficult to access surfaces, see Acculan pre-cleaning and
care information TA016000 (available in Aesculap Extranet at
https://extranet.bbraun.com).
Water qual-
Chemical
ity
D–W
FD–W
Concentrate, alkaline:
– pH ~ 13
– <5 % anionic surfactant
0.5 % working solution
– pH ~ 11*
FD-W
FD-W