The internal datalogger is activated and configured as follows:
Fig. 32:
Activating the internal datalogger
6.4.3

Configuring the outputs

Menu structure:
Fig. 33:
Outputs menu structure
1. Under <SET -> Basic Settings ->, select the item "mem.
interval".
2. Set the scan rate for the analogue signal. The capacity of the
memory is then displayed automatically in the display.
(Example scan rate = 10 sec. -> memory capacity = 170
hours). At a scan rate of 0 seconds (default setting), the ana-
logue signal is not recorded. Pressing the <AUT key> activa-
tes the internal datalogger.
!
Note!
Pressing the <AUT key> irreversibly deletes all the data saved to
the internal datalogger up to that point and starts recording of
the new data.