5 Operation
88
5.6.2
Manual sample volume setting on standard device
Set the required sample volume by shifting the dosing pipe manually.
Calibrate the sample volume as follows:
Fig. 27:
Sample volume calibration
1. Check the set sample volume of the active programme.
2. Loosen the clamping lever and the air tube to remove the
dosing glass.
3. Pull the dosing glass forwards and remove it.
4. Unscrew the bayonet lock and open the dosing glass.
5. Set the sample volume by moving the dosing pipe.
Then put the dosing glass back together and install it in the
reverse order.
ASP Station 2000