For the described functions, a digital signal must be assigned to the digital input 1, 2 or 3. For infor-
mation on the correct wiring of the digital inputs see Chapter 4.1.5.

Analogue input

Menu structure:
Fig. 31:
Analogue input menu structure
Signal
Enter the output signal of the connected device: 0-1 V, 0-20 mA, 4-20 mA
Units
Enter the SI unit of the analogue signal.
Dec. point
Enter the number of decimal places of the analogue signal.
Range
Enter the maximum measuring range.
Note!
For the described functions, an analogue signal must be assigned at the analogue input. For infor-
mation on the correct wiring of the analog input see Chapter 4.1.5.
6.4.2
Configuring the internal datalogger (optional)
As an option, the ASP Station 2000 can possess an internal datalogger. This saves the data of a con-
nected analogue signal and the sampling details (sample volume, bottle filling times, events, etc.).
The analogue value can be a quantity signal, but can also be a pH value, conductivity, turbidity, etc.
Note!
When configuring the analogue input, note the SI unit of the analogue signal.
The minimum scan rate of the internal datalogger is 1 second. You can read off the contents of the
internal datalogger to a PC via the RS-232 interface using ReadWin
5.5 "Communication using ReadWin® 2000").
The input signal stops the selected main programme
The input signal triggers a change to the switching programme.