6 Commissioning
98
Time
Sampling is triggered at settable time intervals.
Quantity
Sampling is triggered after a specific measured flow rate.
!
For this function, a volume signal must be connected to the analogue input or one of the
digital inputs of the sampler. To do this, the digital input must be configured using the
<QUANTITY> function.
Ext. sig.
Sampling is triggered by an external signal.
!
One of the digital inputs must be assigned and configured using the <SAMPLE> func-
tion.
Flow
Sampling is triggered at settable time intervals. The volume sampled is proportional to a
ongoing measured flow.
!
For this function, the ASP Station 2000 must be equipped with a flow-proportional
dosing unit ("twiddle principle"). In addition, a volume signal must be connected to the
analogue input of the sampler.
Shots
Number of samples per sampling cycle.
Distribution
Menu structure:
Program 1
Name
:
Program 1
Sample >
Distribution >
Start-Stop >
Synchronisation >
-mode
Parallel sample >
time
â
á
Esc
<-'
bottle
volume
Esc
Fig. 37:
Distribution menu structure
You can programme a bottle change by time, number or an external signal:
Time
After a set time period has elapsed, the tap switches to the next empty bottle.
Number
After a set number of samples has been taken, the tap switches to the next empty bottle.
Ext. sig.
The tap switches to the next empty bottle when an external signal is present.
!
One digital input must be assigned and configured using the <BOTTLE> function.
Start-stop operation
Menu structure:
Program 1
Name
:
Program 1
Sample >
Distribution >
Start-Stop >
start
Synchronisation >
stop
Parallel sample >
operation
â
á
Esc
<-'
Fig. 38:
Start-stop operation menu structure
Note!
Note!
Note!
Distribution
: time
time
:
00:15 number
:
12
ext.sig.
:
01,0 l
â
á
<-'
Note!
Start-Stop
: time
aut-button
:
prog.end
time
continuous
â
á
Esc
<-'
ASP Station 2000