3 Installation
"
!
76
The flow assembly is used for taking samples from pressurised systems, e.g.:
• Tanks at a height
• Pressure piping
• Conveyance using external pumps
Connections
Fig. 5:
Flow assembly
a
Flow assembly inflow: ¾ "
b
Flow assembly outflow: 1 ¼ "
c
Flow rate: 1000 l/h to 1500 l/h
Caution!
The outlet of the flow assembly must be unpressurised (e.g. drain, open channel).
Application example: sampling from pressure piping
Fig. 6:
Sampling from pressure piping
Use the ball valve V1 to set the flow rate to 1000 l/h...1500 l/h. When the sampling cycle begins,
you can use one of the relay outputs to control and open valve V2. The medium flows through the
pipe and the flow assembly into the outflow. When a settable delay time has elapsed, the sampler
then takes the sample directly from the flow assembly. Valve V2 is closed again once the sample has
been taken.
Note!
Valves V1 and V2 are not contained in the scope of delivery.