6 Commissioning
Creating programmes - Switching programmes
Switch
Sample
Distribution
Parallel sample
Creating programmes - Event programme
Event prog.
Sample
Distribution
Switch
Service
Service
106
Sample: sampling type in the switching programme.
Set
Distribution: distribution mode for the switching programme.
Creating programmes
Switch 1,2
Time: sampling acc. to time.
Set
Volume: sampling acc. to volume (connect volume signal!).
Creating programmes
Ext. sig.: sampling on external signal (connect signal input!).
Switch 1,2
Sample
Flow (optional): sampling acc. to flow (connect volume signal).
Time: bottle change after a specific time.
Set
Number: bottle change after a specific number of samples.
Creating programmes
Ext. sig.:
Switch 1,2
Distribution
Bottle change on external signal (connect signal input!).
Set
Activation and configuration of a sample in separate bottles, in parallel
Creating programmes
to the switching programme.
Switch 1,2
Parallel sample
Sample: sampling type in the event programme.
Set
Distribution: distribution mode for the event programme.
Creating programmes
Event prog.
Time:
Set

Creating programmes
Sampling acc. to time
Quantity:
Event prog.
Sample
Sampling acc. to volume (connect volume signal!).
mode
Ext. sig.:
Sampling on external signal (connect signal input!).
Flow (optional):
Sampling acc. to flow (connect volume signal).
Once:
Single sample.

Time:
Set
Creating programmes
Bottle change after a specific time.
Number:
Event prog.
Distribution
Bottle change after a specific number of samples.
mode
Ext. sig.:
Bottle change on external signal (connect signal input!).
Kein Einf:
Set
Creating programmes
No bottle change for event sample.
Change:
Event prog.
Distribution
Change to the next empty bottle.
Switch
1-x bottles:
Between 1 and x bottles are filled during event sampling.
General:
Set
Service
Software name, software version, software option, CPU number, pre-
set, view angle setting (contrast)
Run times:
Test: Service code necessary!!
Calibration:
Tap calibration (flow-proportional option: sample volume calibration)
6.6

Configuration using Profibus

For a detailed description of the configuration of the ASP Station 2000 using Profibus-DP, please
refer to the "Appendix to the operating manual ASP Station 2000: DP-Slave-Modul_is Pro Gate"
(see 10.10 "Documentation").