Endress+Hauser ASP Station 2000 Mode D'emploi page 104

Masquer les pouces Voir aussi pour ASP Station 2000:
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

6 Commissioning
Sampling
Inputs
Outputs
Memory interval
Thermostat
Interfaces
Creating programmes
Number
Name
Switch (only for an
active switching pro-
gramme)
104
Time set.:
Set
Basic settings
You can change the rinsing, suction, dosing and emptying times manu-
Sampling
ally.
Delay:
You can delay the start of sampling (e.g. after an external signal) by up
to 99 seconds.
CS 1/2:
You can adapt the sensitivity of conductivity probes to different media
(e.g. high for media with poor conductivity).
Dosing:
Support dosing operation with pressure (e.g. for low suction heights) or
make unpressurised.
No flow:
Count and save if there is no inflow during sampling (e.g. with blocka-
ges).
LF2-Igno:
A fault LF2 will be ignored as often as the value is preset (00 to 99).
Sample on ProgStart:
On starting a programme or changing a programme a prompt sample
will be triggered.
Set
Function for digital input 1, 2 and 3 and for setting the analogue input.
Basic settings
Inputs
Set
Function for relay outputs 1, 2 and 3.
Basic settings
Outputs
Set
Time interval for internal recording of measured flow values. 0 sec = no
Basic settings
recording.
Memory interval
Set
Configuration of thermostat in sample compartment; setting of sample
Basic settings
compartment temperature and defrost times.
Thermostat
Set
Transmission rate and definition of the RS232 interface
Basic settings
Interfaces
1: 1 main programme active
Set
2: 1 main programme and 1 switching programme active
Creating programmes
Number
1+Event:
1 main programme and one event programme active
2+Event:
1 main programme, 1 switching programme and 1 event programme
active
Set
Enter names for the four main programmes, two switching programmes
Creating programmes
and the event programme.
Name
Day:
Set
Creating programmes
Switch between main and switching programmes on two specific times
Switch
a day.
Week:
Switch between main and switching programmes on two specific
weekdays.
Q greater:
Switch when flow greater than limit value.
Q smaller:
Switch when flow smaller than limit value.
External:
Switch on external signal.
ASP Station 2000
Automatic
00 sec.
Standard
No press.
Count
00
activated: no
No
Power down
0000 sec
04 °C
1
programme 1-4;
1-U1; 1-E;
1-U1-E
Day
Endress+Hauser

Publicité

Table des Matières
loading

Table des Matières