Trouble-Shooting; Trouble-Shooting Instructions; Process Error Messages - Endress+Hauser ASP Station 2000 Mode D'emploi

Masquer les pouces Voir aussi pour ASP Station 2000:
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

9 Trouble-shooting

110
Order code
Accessory
50031919
Webbed PVC hose 32x5(internal diameter)
50074496
Suction hose inner diameter 13mm, meter
RPS20A-VK
Interface cable for PC, stereo plugs
RPS20A-FB
Bottle tray 6x3 l PE with bottle
RPS20A-FC
Bottle tray 12x1 l PE with bottle
RPS20A-FD
Bottle tray 6x2 l glass with bottle
RPS20A-FE
Bottle tray 12x1 l glass with bottle
RPS20A-FF
Bottle tray 2x12 l PE with bottle
50058741
Bottle 10l PE + lid
51002312
Bottle 12 l PE square incl. lid
51000416
Bottle 20 l PE incl. lid
50088586
Bottle 3 l PE incl. lid
RPS20A-BA
Bottle 1 l PE incl. lid
RPS20A-BB
Bottle 2 l glass incl. lid
RPS20A-B3
Composite container PE 30 l
RPS20A-B6
Composite container PE 60 l
RPS20A-VA
Distribution system (tap, tap drive, distribution frame)
RPS20A-PA
PROFIBUS
51005197
Glass dosing chamber with bayonet connection
50086064
Air filter
51003996
Locking door ASP 2000
9
Trouble-shooting
9.1

Trouble-shooting instructions

Always begin trouble-shooting using the following checklists if faults occur after commissioning or
during operation. Different questions will guide you to the cause of the error and will suggest appro-
priate remedial action.
9.2

Process error messages

Error message
ERROR: RAM
ERROR: Clock
ERROR: EEPROM
ERROR: Conductivity 1
ERROR: Conductivity 2
®
DP slave module
Cause
New programme transmitted
Electronics error
New programme transmitted
EEPROM defective
Contact between cond. 1 - probes
due to water or dirt
Contact at cond. 2 - probe due to
water/dirt
ASP Station 2000
Solution
Acknowledge the message
Repair by E+H Service
Acknowledge message. Repair by
E+H Service.
Clean dosing unit see Chapter 7
Clean dosing unit see Chapter 7
Endress+Hauser