Sampling methods
Q
Flow curve
Q
Time based
Q
Flow based
Q
Flow / time based
Fig. 56:
Sampling types
Sample distribution
The sample liquid is distributed into the individual bottles by means of a tap (item A). In addition to
a 30 l and 60 l composite container, various bottle distributions are available. The distribution ver-
sion can be replaced or changed easily without the need for tools. The ASP Station 2000 allows fle-
xible configuration of the sample distribution. Individual bottles and bottle groups can be freely defi-
ned for the main, switching and event programmes. Individual bottles are located in two separate
bottle trays (item C). Grips on the bottle trays make transportation easy and practical.
Fig. 57:
ASP Station 2000 sample distribution
Item A: tap
Item B: distribution pan
Item C: bottle trays
10 Technical data
Time proportional:
t
A constant sample volume is taken at
constant time intervals.
Quantity proportional:
t
A constant sample volume is taken at
variable time intervals.
t
Flow proportional:
A variable sample volume is taken at
constant time intervals.
t
121