Technical Data; Input; Performance Characteristics - Endress+Hauser COM280 Manuel

Oxymètres de poche
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

COM280
Measured variable
Measuring range
Response time
Reference operating
conditions
Zero current
Maximum measured
2)
error
Long-term drift
Oxygen depletion
(intrinsic consumption)
Electrolyte service life
Temperature
compensation
Salinity correction
2)
In accordance with DIN IEC 746 Part 1, at reference operating conditions
Endress+Hauser
16

Technical data

16.1 Input

Dissolved oxygen [mg/l, % SAT, mbar]
• Measuring range I:
0 to 19.9 mg/l or 0 to 199 %SAT or 0 to 200 mbar
• Measuring range II:
0 to 90 mg/l or 0 to 600 %SAT or 0 to 1250 mbar

16.2 Performance characteristics

From air to nitrogen at 25 °C (77 °F)
t
: 25 s
90
Reference temperature:
Reference pressure:
< 0.1 % of the saturation value
Oxygen concentration, saturation,
partial pressure:
Temperature:
Temperature compensation:
3 % per month
Approx. 6 ng/h at 20 °C (68 °F) and in 1 mg/l dissolved oxygen
Approx. 6 months
0 to 40 °C (32 to 100 °F)
Can be added, fixed salinity 35.0
25 °C (77 °F)
1013 hPa (15 psi)
0.5 % of the measured value ± 1
digit
0.1 K ± 1 digit
< 2 %