Technical Data; Accessories - Endress+Hauser CLS 51 Instructions De Montage Et De Mise En Service

Cellule de mesure de conductivité inductive
Table des Matières

Publicité

Accessoires/Accessories
14
6.

Technical data

Conductivity measurement
Measuring ranges
Lower measuring range limit
Measured value deviation (acc. DIN IEC 746, at 25 °C)
Ambient temperature influence
Ambient temperature
Storage temperature
Relative humidity
Ingress protection (DIN 40050)
Cable introduction
Terminal cross section
Transmitter housing dimensions
Housing material
Weight
Parts in contact with medium
Material of measuring cell
Peak-to-valley height
Thermal conductivity socket
Seal
Medium temperature
Temperature measurement
Measuring ranges
Temperature probe
Measured value deviation (acc. to DIN IEC 746)
Pressure
Tubing cross section required
Dairy fitting, clamp fitting, G 1 1/2"
APV, Varivent connections
Electrical connection
Measuring cable
7.
Accessoires
Voici les accessoires pouvant être comman-
dés séparément :
• OMK
câble prolongateur spécial entre CLS 51 et
Mycom 12-ID / 151 - ID
réf. 50004124
• Joint de boîtier CLS 51
réf. 50073561
0 ... 2000 µS/cm / 20 mS/cm / 200 mS/cm / 1000 mS/cm
± 0,5 % of upper range value
160 x 120 x 70 mm (W x H x D)
aluminium, thermoset resin-coated
O-ring, EPDM (FDA-approved)
sterilisation: 130 °C (max. 30 min)
Pt 100, class B acc. to DIN IEC 751
± 0,5 % of upper range value
7.

Accessories

The following accessories for the CLS 51 may
be ordered separately:
• OMK
Special cable to serve as an extension
between the CLS 51 and the Mycom
121-ID/151-ID
Order no. 50004124
• CLS 51 housing seal
Order no. 50073561
CLS 51
100 µS/cm
± 0,15 %/°C
–10 ... +50 °C
–25 ... +80 °C
5 ... 95 %
IP 67
1 x Pg 16, reduced
2
2.5 mm
2 kg
PEEK
Ra ≤ 0.5 µm
V4A (1.4571)
–5 ... +100 °C
– 35 ... +150 °C
max. 16 bar (20 °C)
min. DN 65
min. DN 50
2
7 x 0.5 mm
, screened

Publicité

Table des Matières
loading