9  Troubleshooting Your STATIM
Problem
Unit does not power ON.
There is water under the machine.
Instruments do not dry.
Solution
Check that the unit is plugged into a properly
grounded outlet and that the power cord is firmly
seated at the rear of the machine.
Try another circuit. Power unit OFF for 10 seconds
and then power ON again.
Check the condition of the line circuit
breaker or fuse.
Check that water was not spilled when refilling the
reservoir. Make sure the plug in the exhaust tube
is secured. Remove and reinsert the cassette.
Attempt another cycle.
Be careful. The metal parts will be hot, and
the cassette will contain hot steam.
The cassette is leaking. If water drips from
the underside of the unit during operation, check
the cassette seal for misalignment or damage and
replace the seal if required.
Be careful. The metal parts will be hot, and
the cassette will contain hot steam.
Attempt another cycle. If it still leaks attempt another
cycle using a different cassette if possible.
If the leak persists, turn the unit OFF, remove and
unload the cassette, unplug the unit, and call
your dealer.
Best drying occurs when the cycle continues to comple-
tion. Allow the cycle to finish. Make sure the instruments
are loaded correctly in the cassette. Refer to section 4.
Using Cassettes and Preparing Instruments.
Check the unit leveling.
Check air/biological filters and replace if dirty. Clean
the inside of the cassette and treat with STAT-DRI Plus
drying agent. Refer to section 8.1 Cleaning the
Cassette. Examine the exhaust tube (tube to the
waste bottle) for kinks.
Page 40