5  Using your STATIM
5.1.2 Hollow Wrapped Cycles
The STATIM 2000 G4 and 5000 G4 feature two WRAPPED 134°C type S sterilization
cycles.These cycles are used to sterilize solid and hollow metal instruments which have been
sealed in paper/paper or paper/plastic autoclave bags.
From the main menu, press the WRAPPED icon to scroll through the available cycles.
HOLLOW WRAPPED
Press START
Once the desired cycle has been selected, press the START button.
The STATIM G4 will remember the last Hollow Wrapped cycle type selected and will display
it when you select the Hollow Wrapped icon.
NOTE: A Process Challenge Device (PCD) is available for validation of HOLLOW
WRAPPED (S) 134°C/3.5 min cycle.
HOLLOW WRAPPED (S) 134˚C / 3.5 min
HOLLOW WRAPPED (S) 134˚C / 18 min
HOLLOW WRAPPED (S) 134˚C / 3.5 min
HOLLOW WRAPPED (S) 134˚C / 18 min
Page 26
HOLLOW WRAPPED
Press START
A
B
C
A
100˚C / 101 kPa
B
6 purges (Conditioning)
peak at 133˚C / 295 kPa
vent to 115˚C / 169 kPa
C
Pressurizing
D
E
F
3.5
min /
18 min
D
Sterilizing
134˚C / 304 kPa -
138˚C / 341 kPa
E
Venting
F
Air Drying