Stopping A Cycle - SciCan AUTOCLAVE Statim 2000 G4 Manuel De L'utilisateur

Masquer les pouces Voir aussi pour AUTOCLAVE Statim 2000 G4:
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

5  Using your STATIM
5.2  Running a Cycle 
cont'd
The buzzing noise during the air drying stage is
the compressor operating. The air drying phase of
the cycle may be interrupted at any time by pressing
the STOP button.
When the automatic 60-minute air drying stage is finished
and the sterilization cycle is successful, the touchscreen
will display a Cycle Complete message and the reminder
tone will sound until the STOP button is pressed
or the cassette is removed from the unit.
Be careful. The metal parts will be hot and the cassette
may contain hot steam.
5.3  Stopping a Cycle
To stop a cycle, press the STOP icon at the bottom right
of the touchscreen. If the STOP button is pushed,
the cassette is removed, or the unit detects a problem
while operating, the cycle will stop. Once a cycle has been
stopped, the STOP button must be pressed before another
cycle can be started. The display will show:
If the display shows the message, CYCLE FAULT or NOT STERILE, the cassette contents
are not sterile! See Section 9. Troubleshooting for more information.
If the air drying stage of the cycle is interrupted, do not store wrapped instruments that were
in the cassette unless they are dry.
NOTE: If Process Enforced usage is enabled, a PIN screen
will appear after you have pressed STOP. To release the load
manually, press EN on the PIN screen. On the next screen,
select RELEASE LOAD MANUALLY.