9  Troubleshooting 
Problem
Excessive steam issuing from the front
of the machine.
Machine will not start and touchscreen shows:
Machine will not start and touchscreen shows:
The printer does not work.
Time and date are incorrect.
con't.
Solution
Remove and reinsert the cassette. Attempt another
cycle. Remove and check the cassette seal for
misalignment or damage. Replace the seal if
required. Be careful as the metal parts will be hot and
the cassette will contain hot steam.
If the leak persists, turn the unit OFF, remove and
unload the cassette and contact your SciCan dealer.
Press on the icon to confirm whether it is a water level
problem or a water quality problem.
If it is a water quality problem, you have likely used
water which is not steam-process distilled or is
improperly distilled.
Empty the reservoir and refill with steam-process dis-
tilled water containing less than 5 ppm total disolved
solids (having conductivity of less than 10 µS / cm).
If you have the water conductivity meter, check the
quality of the water before refilling the reservoir.
To empty the reservoir, see section 8.11 Shipping the
unit / Draining the Reservoir.
Press on the icon to confirm whether it is a water level
problem or a water quality problem.
If the level of the water in the reservoir is low, refill the
reservoir. Refer to the steps described in section 3.4
Filling the Reservoir.
Make sure that the printer cable is connected securely
with the connector on the back of the STATIM. Make
sure that the printer is powered ON. Power unit OFF
for 10 seconds and then power ON again.
The time and date have not been set. See Section 3.
Setting up your STATIM, for time and date instructions.
Page 42