4  Using Cassettes and Preparing Instruments
Wrapped Instruments (single wraps)
Place the instruments into single layer autoclave bags according to the manufacturer's
instructions. Orient the instrument rack in the cassette to ensure that wrapped instruments
rest approximately 6 mm / 0.25" above the cassette base. Place the wrapped instruments
on the rack and arrange them to avoid overlap. Ensure that all wrapped loads are dry before
handling and /or storage to maintain sterility.
For STATIM 2000 G4:
The use of cloth wraps in the STATIM is not recommended.
SciCan recommends the use of paper / paper and plastic / paper autoclave bags
manufactured in accordance to EN 868. Loosely pack instruments in the bags to allow
steam penetration to all instrument surfaces.
For STATIM 5000 G4:
The unwrapped instrument rack fitted with a maximum of 10 drying plates will hold 10 auto-
clave bags. Care must be taken to ensure that the combined weight of the loaded bags does
not exceed 1.5 kg (3.3 lbs).
Rubber and Plastic Instruments
The following materials can be sterilized in the STATIM:
Nylon, polycarbonate (Lexan™), polypropylene, PTFE (Teflon™), acetal (Delrin™),
polysulfone (Udel™), polyetherimide (Ultem™), silicone rubber, and polyester.
When loading rubber and plastic instruments in the tray, leave a space between the
instruments and the cassette walls. This ensures that steam reaches all surfaces, and
will promote drying.