8  Maintaining your STATIM
8.8  Maintaining Fluid Levels
1. The water reservoir level is continually monitored by your STATIM. If the reservoir is low,
a red X will appear on the water icon        of the cycle select screen. Press the icon to go
to the next screen to confirm that it is a water level issue and not a water quality issue.
2. If the reservoir is low, a        will appear next to WATER LEVEL.
3. To fill the reservoir, remove the cap from the top of the unit and fill the reservoir.
We recommend using a funnel to minimize spills. Each time you refill the reservoir,
empty the waste bottle and refill with water to the MIN line. Empty the waste bottle often
to avoid unpleasant odors and discoloration of the contents. (A low-level chlorine-free
disinfectant, prepared according to the manufacturer's instructions, may be added
to the waste bottle to remedy this situation).
8.9  Reading Water Quality
1. The water quality is continually monitored by your STATIM. Use only steam-process
distilled water containing less than 5 ppm total dissolved solids (having conductivity of
less than 10 μS / cm) in the STATIM. If the water quality is beyond these tolerances,
a red X will appear on the water icon        of the cycle select screen. Press the icon to go
to the next screen to confirm that it is a water quality issue and not a water level issue.
2. If the water quality not suitable,        will appear next to micro S. and parts per million
values.
3. Using the drain tube (see section 3.5 Priming the Pump, Figure 6) empty the contents
of the reservoir into the water container and replace with steam-process distilled water
containing less than 5 ppm total dissolved solids (having conductivity of less than
10 μS / cm).