8  Maintaining your STATIM
8.10  Preventative Maintenance Schedules
To ensure trouble-free performance, both the operator and the dealer must follow
a preventative maintenance schedule.
NOTE: Please refer to your National, Regional, State or Safety laws for any additional reoccurring
user testing that may be required.
The schedules below describe the necessary actions.
Water Reservoir
Waste Bottle
Cassette
Water Filter
Biological and/or
Air Filter
Air Filter
Biological Air Filter
Cassette Seal
Cassette
Biological Filter
Solenoid Valve
Pump
Check Valve
Water Reservoir
Calibration
Operator
• Drain after each working day.
• Empty the waste bottle every time you refill the reservoir. Fill with tap water up to MIN line.
• You can also add some chlorine-free disinfectant.
• Wash the interior of the cassette with chlorine-free dishwashing detergent or soap.
Rinse thoroughly with water to remove all traces of the detergent and treat
interior surfaces of the cassette with the STAT-DRI™ Plus drying agent to enhance
the drying process.
Order more STAT-DRI ™ Plus from SciCan quoting 2OZPLUS, 8OZPLUST,
or 32OZPLUS.
• Check the water reservoir filter every week and clean if necessary. Replace only if necessary.
• Check the filter for dirt and moisture. Replace if dirty. Call for service if wet.
• Replace every 500 cycles or six months (whichever comes first).
• Replace every 500 cycles or six months (whichever comes first).
• Replace every 500 cycles or six months (whichever comes first).
Technician
• Check the tray, lid and seal for damage. Replace if necessary.
• Inspect the biological filter for moisture.
• Inspect the valve and clean if dirty. Replace the plunger if defective.
• Clean the filters, replace if dirty.
• Remove the exhaust tube from the back of the unit during the air drying phase. Check for
air coming from the fitting.
• Remove the air compressor tube from the check valve inlet while running a cycle. Make
sure no steam is leaking from the valve. Replace if there are any leaks.
• Check the reservoir for dirt. Clean and rinse with steam process distilled water if necessary.
• Calibrate the unit.
Page 38