5  Using your STATIM
5.2  Running a Cycle
To operate each cycle, follow these steps.
1. Turn the power switch at the back of the unit to ON.
At start up, the unit will display the main menu.
mm/dd/yyyy
SELECT A CYCLE
2. Press the appropriate cycle button on the touchscreen to scroll through the
available cycles.
The display will show the cycle name and parameters. The STATIM G4 features
seven different cycles:
HOLLOW UNWRAPPED
(S) 134˚C / 3.5 min
HOLLOW UNWRAPPED
(S) 134˚C / 18 min
SOLID UNWRAPPED (N)
134˚C / 3.5 min
3. When you have found the cycle you require, press the START icon.
HOLLOW WRAPPED
Press START
NOTE: If Process Enforced usage is enabled, a PIN screen will appear after you have
pressed START. Enter your PIN to start the cycle.
hh:mm:ss
HOLLOW WRAPPED (S)
or
134˚C / 3.5 min
HOLLOW WRAPPED (S)
134˚C / 18 min
or
HOLLOW WRAPPED
Press START
Page 29
RUBBER / PLASTIC (S)
121˚C / 15 min
RUBBER / PLASTIC (S)
121˚C / 30 min