8  Maintaining your STATIM
8.1  Cleaning the Cassette
Keeping the STATIM cassette clean is good clinical practice and assists in the proper
functioning of the unit. SciCan recommends that the interior surface be cleaned at least once
a week. Cleaning the inside of your cassette is very important if you regularly sterilize
lubricated instruments.
1. Use dishwashing soap or a mild detergent that does not contain chlorine.
2. Scrub the inside of the cassette with a cleaning pad designed for use with Teflon™ coated surfaces.
3. After scouring, rinse thoroughly with water to remove all traces of the detergent.
Coating the entire inside surface with STAT-DRI Plus drying agent induces water to form an
even coat on the inside surface, without beading. The water in contact with the hot cassette
surfaces also evaporates much more efficiently. Spotting is minimized and instruments dry much
better. STAT-DRI Plus should be applied every 10 cycles, and after every cassette cleaning.
STAT-DRI Plus is available from SciCan in a 2-ounce bottle (SciCan Part # 2OZPLUS) or
an 8-ounce bottle (SciCan Part #8OZPLUS) or a 32-ounce bottle (SciCan Part #32OZPLUS).
8.2  Cleaning the Water Reservoir Filter
This water reservoir filter should be cleaned at least once a week or when required.
The filter can easily be removed and cleaned by placing the filter upside down under running
water to wash away the particles until clean, and then placed back into the reservoir opening.
If a replacement water reservoir filter is required, order part number 01-109300S.
8.3  Cleaning the Reservoir
Check the reservoir for dirt or particles. The reservoir may be cleaned by draining followed
by cleaning and rinsing with steam process distilled water ONLY. Use of chemicals or
cleaning agents is not recommended and could damage the unit.
8.4  Cleaning the Exterior Surfaces
Use a soft cloth moistened with soap and water to clean all exterior surfaces.
Do not use harsh cleaning chemicals or disinfectants.
8.5  Changing the STATIM 2000 G4 Air Filter
The filter should be replaced every six months or after 500 cycles in order to maintain an
adequate supply of clean air during the air drying cycle.
To change the filter, follow these steps:
1. Turn the power switch at the back of the unit OFF.
2. Remove and discard the old foam air filter
3. Install the new filter
4. Secure the filter plate
the compressor using the screw
during the disassembly procedure.
n
3
4
(SciCan part no. 01-100207S).
n
2
2
to the back of
n
1
1
retained
Page 34
2000 G4
n
3
4
.
1
3
2
Figure 10