Size: 7 x 9 in (178 x 229 mm)
Prompt
Reservoir weight not detected. Self-start
not available. Press
FCT 1: Empty the bowl to the reservoir.
Press and hold
to run the pump. Press
when complete.
FCT 2: Empty the bowl to the holding bag.
Press and hold
to run the pump. Press
when complete.
FCT 3: Invert the holding bag. Press and
hold
to remove air. Press
plete.
Vacuum pressure exceeded the setting.
Make sure the suction line is not clam-
ped. Press
to resume.
a
Prompt will have different audible and visual alert signals when issued during setup.
18.5 Error prompts, audible and visual alert signals
Error prompts may appear during device operation or startup.
Error
The vacuum system failed. Vacuum is not
available. Press
18.6 Irrecoverable errors, audible and visual alert signals
These errors cannot be recovered by the device or user. The device should be taken out of service until it is examined by a Medtronic
service representative or biomedical equipment technician trained and certified by Medtronic.
Error
Failure of any of the following components
will produce a message containing the com-
ponent name, the error code, and a state-
ment to contact Medtronic for service:
• Air detector
• Centrifuge
M967226A006 1B Medtronic Confidential
Response
10. Turn off the device, then turn it on.
11. If the previous step does not resolve the mes-
sage, use another device if available. Contact
your Medtronic service representative.
Warning: Close all clamps and use caution if
transferring the disposables to a new device.
The weight sensor does not detect the reservoir.
to manually start.
1. Mount the reservoir in the reservoir holder
(enabling self-start) if self-start mode is
desired.
2. Or, press
FCT 1 was selected while the device was in Stop
mode. These functions enable the pump to be man-
ually engaged.
FCT 2 was selected while the device was in Stop
mode. These functions enable the pump to be man-
ually engaged.
FCT 3 was selected while the device was in Stop
when com-
mode. These functions enable the pump to be man-
ually engaged.
This message appears if the device vacuum system
could not regulate the vacuum pressure, possibly
because of a clamp on the vacuum line between the
fluid trap and the reservoir.
1. Confirm that vacuum and suction lines are not
clamped or immersed in liquid.
2. To continue, press .
3. Press
4. If the previous steps do not resolve the issue,
turn off the power, then turn it on, and retry.
Note: If prompt returns, the device can still be used
if an external regulated vacuum source is connected
to the reservoir. Contact your Medtronic service
representative.
Response
This message appears if the device vacuum system
to resume.
has failed.
1. Confirm that vacuum and suction lines are not
clamped or immersed in liquid.
2. Turn off the power, then turn it on, and retry.
3. If the previous steps do not resolve the issue,
contact your Medtronic service representative.
Note: The device can still be used if an external
regulated vacuum source is connected to the
reservoir.
Response
Contact your Medtronic service representa-
tive, and provide the error code displayed on
the screen.
Warning: Close all clamps and use caution if
transferring the disposables to a new device.
to manually start.
to restore vacuum.
Audible alert
signal
Medium
47
Audible alert
Visual alert
signal
signal
Notify
None
None
None
None
None
None
None
Medium
Medium
Audible alert
Visual alert
signal
signal
Medium
Medium
Visual alert
signal
Medium
English
Composed: 2018-06-08 06:06:55
PSJ - Size-selectable package manual 02-MAY-2018
502431-132