Figure 35. Emergency setup
1 Anticoagulant solution containers
2 Suction/anticoagulant lines to patient
3 Blood collection reservoir
1. Collect additional supplies necessary for this configuration:
• Suction/anticoagulant line
• Adequate quantity of anticoagulant solution
2. Attach the vacuum line to the yellow-capped port on the collection reservoir.
3. Attach the other end of the vacuum line to a regulated vacuum source.
4. Attach the blue-capped connector of the additional suction/anticoagulant line to one of the available inlet ports on the collection
reservoir.
5. Close the roller clamp on the suction/anticoagulant line. If the anticoagulant container is nonvented, spike the anticoagulant
container and open the vent cover on the drip chamber.
6. Spike the additional anticoagulant container for the second suction/anticoagulant line.
17.2 Using 2 reservoirs in series
Note: If necessary to transition to 2 collection reservoirs midcase, follow the steps in this section.
English
M967226A006 1B Medtronic Confidential
4 Step-down connector to device
5 Vacuum line to regulated vacuum source
38
502431-132
Composed: 2018-06-08 06:06:55