Prompt
Holding bag may be full. Empty or change
if necessary. Press
Machine paused. Press
Machine stopped. Press
start.
Self-start enabled. Press
start.
The pump is running too slow. Check the
kit installation. Press
Centrifuge is running too slow. Check the
bowl installation. Press
English
M967226A006 1B Medtronic Confidential
Response
1. Examine the holding bag line for closed clamps
and other occlusions.
2. Examine the holding bag. Empty or replace the
holding bag if it is full.
3. Clean the air detector and tubing with water,
and then dry.
4. If the message is not resolved, use another
device if available. Contact your Medtronic
service representative.
Warning: Close all clamps and use caution if
transferring the disposables to a new device.
The device estimates the volume pumped to the
to resume.
holding bag and a notice appears when the bag is
full.
1. Empty or replace the holding bag.
2. Press .
to resume.
The pump has been manually paused using .
1. Resolve the issue.
2. Press
to resume.
to enable self-
This message appears after the disposable kit is
installed. Enable self-start by pressing . The
device automatically starts when it detects approx-
imately 800 mL of blood in the reservoir for at least
5 s.
to manually
Press
to cause a manual start.
Tubing is binding on the pump.
to resume.
1. Confirm that the pump is moving. If it is moving,
proceed to the next step. If not, contact your
Medtronic service representative.
2. Ensure that the tubing is correctly inserted in
the pump tubing channel.
3. If the previous steps do not resolve the mes-
sage, remove the tubing from the pump tubing
channel.
4. Remove the pump head.
5. Confirm that the pump head rollers move
freely, and are free of burrs and other debris.
6. Clean the pump head and rollers if necessary
using an approved cleaning solution (Chap-
ter 16).
7. Reinstall the pump head.
8. If the previous steps do not resolve the mes-
sage, use another device if available. Contact
your Medtronic service representative.
Warning: Close all clamps and use caution if
transferring the disposables to a new device.
The bowl may be binding in the centrifuge.
to resume.
1. Unlock then relock the centrifuge bowl to
ensure that it is correctly installed.
2. Remove the centrifuge bowl.
3. Examine the bowl for damage.
4. Discard and replace the bowl if it is damaged.
5. Examine the centrifuge chamber for debris and
liquid.
6. Examine the centrifuge chamber for binding by
moving the chamber by hand. If it is binding,
contact your Medtronic service representative.
7. Remove the debris or liquid if necessary.
8. Reinstall the bowl and confirm normal opera-
tion.
9. If the previous steps do not resolve the mes-
sage, document all fluid volumes.
Audible alert
signal
Medium
None
None
None
Medium
Medium
46
502431-132
Visual alert
signal
Medium
Low
None
None
Medium
Medium