11. Before blood collection, open the roller clamp and prime the secondary collection reservoir with a minimum of 200 mL of
anticoagulant solution.
12. Reduce the anticoagulant flow to each collection reservoir to a ratio of approximately 15 mL of solution to 100 mL of blood.
13. Monitor both collection reservoirs periodically for appropriate anticoagulation.
14. To move blood from the secondary collection reservoir to the primary collection reservoir, open the clamp on the connector to
the primary collection reservoir, and momentarily clamp the suction/anticoagulant line on the secondary collection reservoir,
using a hemostat.
15. To pause the movement of blood from the secondary collection reservoir to the primary collection reservoir, open the hemostat
on the suction/anticoagulant line to the secondary collection reservoir, and close the clamp on the connector to the primary
collection reservoir.
16. To maintain the desired vacuum setting during processing using 1 vacuum source, either clamp one of the
suction/anticoagulant lines not in use, or increase the vacuum setting to maximum.
17.3 Using 2 reservoirs in parallel
Figure 37. Using 2 collection reservoirs in parallel
1 Anticoagulant solution
2 Suction/anticoagulant line to patient
3 Primary blood collection reservoir
4 Tandem reservoir Y assembly
1. Collect additional supplies necessary for this configuration:
• Blood collection reservoirs
• Suction/anticoagulant line
• Adequate quantity of anticoagulant solution
• Tandem reservoir Y assembly
• Vacuum source, if necessary
• Vacuum line, if necessary
• 1 reservoir holder: either the type that attaches to the device rear reservoir holder mounting base (Figure 6) or the
Medtronic cardiotomy reservoir holder E302, which attaches to an external IV pole.
Warning: Do not attach a reservoir to the device IV pole using a reservoir holder such as the Medtronic E302. The weight of
a filled reservoir attached to the device IV pole may cause tipping.
2. Attach a second reservoir holder to the device rear reservoir holder mounting base or to an external IV pole.
3. Place the secondary collection reservoir into the secondary reservoir holder.
4. Do not attach the step-down connector to either collection reservoir. Instead, use a tandem reservoir Y assembly to connect
the primary and secondary collection reservoirs. Connect the single tubing to the secondary collection reservoir and the Y
assembly to the primary collection reservoir. Ensure that the connectors are fully secured.
English
M967226A006 1B Medtronic Confidential
5 Secondary blood collection reservoir
6 Vacuum line to the vacuum overflow trap of the device
7 Vacuum line to external regulated vacuum source