1 Current case data area
2 Screen selection area
Current case data area
• Fill: The volume of blood transferred from the collection reservoir to the centrifuge bowl in the current processing cycle
• Wash: The volume of saline used for the current processing cycle
• Empty: The volume of washed red blood cells transferred to the holding bag in the current processing cycle
• Bowl (X): The number of completed Fill, Wash, and Empty processing cycles
• Total: The total Fill, Wash, and Empty volumes for a case
Screen selection area
• Main: Contains current case data, user controls, and prompts
• Data: Allows the user access to data stored in the device (Chapter 14)
• Settings: Allows the user to adjust the device settings (Chapter 12)
: Allows the user to enter patient ID, operator ID, and disposable lot, heparin lot, and ACD-A lot data (Section 13.2)
: Allows the user to access the Service and maintenance screen (Section 19.3)
Note: When
Section 19.4.
User controls area
• Vacuum: Allows the user to set the vacuum setting. The vacuum setting can also be adjusted during a case. The default and
maximum vacuum settings can be preset using the Settings screen (Section 12.1).
• Wash volume: The default wash volume is 250 mL. The wash volume can also be adjusted during a case; adjustments are
effective immediately. If another volume is desired, select Fast or Emergency from the Wash volume menu on the Main
screen. Fast uses 100 mL of wash solution. Emergency does not use any wash solution.
Note: Selecting Emergency causes a notification to appear every 10 min. If the notification is postponed, the notification will
reappear at the interval selected in the Reset alert field. See Section 19.4.
: Allows the user to begin or resume automated blood processing, and to control the manual FCT functions.
during operation; press
Note: If the centrifuge is operating when
automatically returns the bowl contents to the reservoir.
• Function: Allows the user to manually transfer volumes or remove air from the holding bag (Figure 21).
: Allows the user to mute and unmute the audible signals. When muted, the button icon changes, providing a visual indication
of the selection.
Note: The device automatically reverts to the unmuted state after 60 s.
• Date & Time: Displays the local date and time. The flashing : indicates that the user interface is operating normally.
Note: The date and time can be adjusted on the Settings screen (Chapter 12).
Prompts area
The Prompts area presents prompts and notifications. Prompts are messages that require user action. Notifications provide
information about device conditions that are important but do not stop device operation. See Section 18.4 and Section 19.4 for
prompts and notifications that may appear during operation.
12.7 Function capabilities
Note: Function capabilities are only enabled when the device is in Stop mode.
1. Press
while the pump is running, to empty the bowl to the reservoir. Wait for the bowl to empty completely, or press
to immediately enable all Function capabilities.
English
M967226A006 1B Medtronic Confidential
is highlighted in yellow, it indicates the presence of a notification. Click
to pause the applicable operation.
is pressed, the paused state persists for a maximum of 60 s. After 60 s, the device
3 User controls area
4 Prompts area
to view the notification. See
24
502431-132
changes to
again