Component/area
Centrifuge
Pump head
Vacuum overflow trap
English
M967226A006 1B Medtronic Confidential
Cleaning
Note: To clean blood spills, use a 10% bleach solution or equivalent disinfectant solution.
2. Carefully wipe across the air detector or valve assembly to remove any blood, dust, or debris.
3. Carefully wipe across the air detector or valve assembly with another foam-tipped swab to
remove any remaining moisture.
1. If blood or fluids spill in the centrifuge chamber, turn off the device.
2. Remove the wash kit.
3. To collect disinfectant solutions during the cleaning process, slide the empty tray from the
wash kit into the wash kit tray holder under the device, from the right side. Tilt the tray slightly
upward in order to insert the edges of the tray into the wash kit tray holder.
4. Confirm that the wash kit tray is correctly installed.
Warning: While cleaning the centrifuge, failure to correctly install the tray for fluid collection will
lead to spillage.
Note: The centrifuge drain is equipped with female pipe threading (1/4-18 NPT). If desired, a
customized fluid container can be used to collect fluids, instead of using the wash kit tray.
5. Turn on the device.
6. Press
on the touchscreen to access the Service and maintenance screen.
7. Press Clean centrifuge.
8. Press
to start the centrifuge rotating.
Warning: Failure to press
will result in the centrifuge not rotating and a failure to effectively
clean the centrifuge.
9. Over a 1 min to 2 min period, slowly pour 100 mL of a 10% bleach solution or equivalent
disinfectant solution into the centrifuge chamber while it is rotating until it starts to overflow.
During rotation, fluid will spin out of the centrifuge and into the tray. Continue adding solution
until the centrifuge is clean.
Warning: While cleaning the centrifuge, failure to add sufficient volume of a 10% bleach
solution or equivalent disinfectant solution for a sufficient duration can lead to ineffective
cleaning.
10. Over a 1 min to 2 min period, rinse the centrifuge by slowly pouring 100 mL of water into the
centrifuge chamber.
11. Press .
Caution: Failure to press
can lead to minor injury if the chamber is contacted while it is
spinning.
12. Dispose of the used cleaning solutions in accordance with hospital procedures for the dis-
posal of biohazardous waste.
13. Dry the chamber with a soft cloth. Do not use alcohol to clean the centrifuge chamber. Remove
and empty the tray.
Clean the exterior and interior of the pump head after every use. To clean the pump head, perform
the following procedure:
1. Turn off the device and unplug the power cord from the supply mains.
2. Pull back the pump lever with one hand.
Caution: Be careful not to injure fingers in the pump lever.
3. With the other hand, lift the pump head up and off of the shaft.
4. Release the pump lever.
5. Clean the spilled fluids from the pump head as much as possible, using towels and foam-
tipped swabs dampened with a 10% bleach solution or equivalent disinfectant solution. Dry
the area thoroughly.
Caution: Failure to thoroughly dry the area can result in corrosion.
6. Ensure that the pump head rollers move freely. Clean the pump head rollers with soap and
water or with water alone. Dry the area thoroughly.
7. To reinstall the pump head onto the shaft, pull back the pump lever with one hand.
8. With the other hand, rotate the pump head while applying pressure until the keyed slot in the
pump head aligns with the shaft and the pump head begins to slide onto the shaft.
Note: The dimple on the pump head aligns with the shaft when the pump head slides onto the
shaft.
9. Release the pump lever.
1. After every use, examine the vacuum overflow trap for fluids or contamination.
2. If fluids are present or if the vacuum overflow trap is dirty, remove, disassemble, and clean the
vacuum overflow trap according to approved hospital protocols with a 10% bleach solution or
equivalent disinfectant solution.
3. Rinse with water and allow the components to dry prior to reassembly and reconnection to the
device.
36
502431-132
Composed: 2018-06-08 06:06:55