13. Confirm that there are no kinks in the waste line.
Note: If desired, the wash kit tray may be removed and the cradle arm rotated to its original position.
14. Open the manifold cover.
Note: See the top panel components in Figure 3 and Figure 4.
15. Place the manifold into the recessed cavity.
16. Place the 3 manifold tubing lines into the tubing guides, and ensure that each tube is fully seated in each guide.
17. Place the single tubing line of the manifold into the air detector, and ensure that it is fully seated.
18. Close the manifold cover.
Note: There is a feature on the manifold cover that pushes the tubing into the air detector when closing the cover.
19. Rotate the manifold cover latch until the lock indicators are aligned.
20. Pull the pump lever and hold it open.
21. Stretch the pump tubing around the pump head and over the pump outlet tubing guide. Release the pump lever.
22. Close the saline line clamps.
23. Spike the middle manifold tubing into the saline bag or bags.
Note: If using only 1 saline bag, confirm that the clamp on the unused saline line is closed.
24. Open the saline clamp or clamps.
25. Connect the left-hand tube of the manifold to the blue-capped step-down connector at the bottom of the reservoir, and then
open the clamp on the step-down connector.
26. Respond to prompts on the touchscreen.
Table 1. Setup of disposables prompts, audible alert signals, and visual alert signals
Prompt
Kit is not installed. Lock the bowl
cover.
Insert the kit in the valve and the pump.
Press
to resume.
Kit is not installed. Lock the manifold
cover. Press
to resume.
M967226A006 1B Medtronic Confidential
Response
The sensor has detected that the bowl cover is not
locked.
1. Place the bowl in the centrifuge chamber with the
yellow-capped waste line outlet tubing facing the
waste bag. Refer to the markings on the top
panel.
2. Align the centrifuge markings with the vertical
assemblies on the bowl.
3. To lock the bowl, simultaneously press down on
the bowl and turn it clockwise until a click is heard.
The kit is not properly installed in either the manifold
a
valve or the pump.
1. Open the manifold cover.
2. Place the manifold into the recessed cavity.
3. Place the 3 manifold tubing lines into the tubing
guides, and ensure that each tube is fully seated
in each guide.
4. Place the single tubing line of the manifold into
the air detector, and ensure that it is fully seated.
5. Close the manifold cover.
Note: There is a feature on the manifold cover
that pushes the tubing into the air detector when
closing the cover.
6. Rotate the manifold cover latch until the lock indi-
cators are aligned.
7. Pull the pump lever and hold it open.
8. Stretch the pump tubing around the pump head
and over the pump outlet tubing guide.
9. Release the pump lever.
A sensor has detected that the manifold cover is not
a
locked.
1. Open the manifold cover.
2. Place the manifold into the recessed cavity.
3. Place the 3 manifold tubing lines into the tubing
guides, and ensure that each tube is fully seated
in each guide.
4. Place the single tubing line of the manifold into
the air detector, and ensure that it is fully seated.
5. Close the manifold cover.
Audible alert
signal
None
None
None
29
502431-132
Visual alert
signal
None
None
None
English
Composed: 2018-06-08 06:06:55
PSJ - Size-selectable package manual 02-MAY-2018