Medtronic autoLog IQ Mode D'emploi page 39

Système d'autotransfusion

Publicité

Les langues disponibles
  • FR

Les langues disponibles

Size: 7 x 9 in (178 x 229 mm)
Figure 36. Using 2 collection reservoirs in series
1 Anticoagulant solution
2 Suction/anticoagulant line to patient
3 Primary blood collection reservoir
4 Step-down connector to device
5 Secondary blood collection reservoir
1. Collect additional supplies necessary for this configuration:
• Blood collection reservoirs
• Suction/anticoagulant line
• Adequate quantity of anticoagulant solution
• Vacuum source, if necessary
• Vacuum line, if necessary
• Medtronic cardiotomy reservoir holder E302, which attaches to an external IV pole
Warning: Do not attach a reservoir to the device IV pole using a reservoir holder such as the Medtronic E302. The weight
of a filled reservoir attached to the device IV pole may cause tipping.
2. Attach a second reservoir holder to an external IV pole, above the primary collection reservoir.
3. Attach a step-down connector to the bottom of the secondary collection reservoir. Ensure that the connector is fully secured.
4. Place the secondary collection reservoir into the holder above the primary reservoir holder.
5. Connect the step-down connector from the secondary collection reservoir to an available inlet port on top of the primary
collection reservoir, and close the clamp on the step-down connector.
6. If using 1 regulated vacuum source, connect tubing from a dual vacuum assembly to the yellow-capped ports on each
collection reservoir. Connect the single tubing from the dual vacuum assembly to the vacuum source.
Note: The dual vacuum assembly is a "T" shaped tubing component, included with the tandem reservoir Y assembly, which
connects 1 vacuum pump to the vacuum ports of the 2 reservoirs.
7. If using a secondary regulated vacuum source, attach the second vacuum line to the yellow-capped port on the secondary
collection reservoir. Attach the other end of the second vacuum line to the secondary regulated vacuum source.
8. Attach the blue-capped connector of an additional suction/anticoagulant line to one of the available inlet ports of the secondary
collection reservoir.
9. Close the roller clamp on the additional suction/anticoagulant line. If the anticoagulant container is nonvented, spike the
anticoagulant container and open the vent cover on the drip chamber.
10. Spike an additional anticoagulant container for the second suction/anticoagulant line.
M967226A006 1B Medtronic Confidential
6 Step-down connector from secondary collection reservoir to
primary collection reservoir inlet port
7 Vacuum line to regulated vacuum source
8 Vacuum line to device
39
502431-132
English