Installation
• The direction of flow is marked on the column.
• 1.8 µm columns (ZORBAX RRHT, ZORBAX RRHD) can only be
operated in the flow direction marked on the column.
• For removable, zero-dead volume column connections Agilent
recommends the use of the InfinityLab Fittings. Choices are:
Maximum
System Pressure
Up to 400 bar
Up to 800 bar
Up to 1300 bar
For more information and part numbers, please see to the Agilent InfinityLab Fitting
Brochure (5991-5164EN).
InfinityLab Quick Connect
assembly, p/n 5067-5961
Learn more at www.agilent.com/chem/infinitylabfittings
Additional care must be taken to avoid damage to PEEK or PEEK-lined
columns during installation. Combined compression and rotation may cause
damage, therefore fittings without ferrules such as PEEK Finger-tight Fittings
are not recommended. Fittings should not be overtightened, and should be
swaged to the correct depth to avoid damage.
Recommended Fitting
InfinityLab Quick Turn Fitting
(finger-tight)
InfinityLab Quick Turn Fitting
(with Mounting tool)
InfinityLab Quick Connect Fitting Fitting: 5067-5965
ENGLISH
Part Numbers
Fitting: 5067-5966
Fitting: 5067-5966
Mounting Tool: 5043-0915
InfinityLab Quick Turn
fitting, p/n 5067-5966