WARNING: Do not over-tighten. Overtightening can
damage the bone surface.
Sterile Latarjet Screw for Fixation
Verify the lumen length by using the supplied reference
instrument.
CAUTION: Correct selection of the screw length is
important.
• For correct positioning of the graft and the implant, first
insert the Glenoid K-Wire (provided in the Instrument Set) to
the required location and introduce the Sterile Latarjet Screw
over the wire.
• Tighten the Sterile Latarjet Screw (with or without) the
bushing.
WARNING: Care should be exercised; over-tightening to
high torque levels can cause damage to the bone and may
lead to compromised holding strength.

HOW SUPPLIED

Devices are provided STERILE. Re-use and/or reprocessing
are prohibited.
Discard after procedure, even if not used.