Illustration and Nomenclature
1. Jaws
2. Atraumatic teeth
3. I-BLADE® knife (not visible)
4. Shaft
5. Rotation Knob
Transport and Storage Conditions
Temperature: -20° C to 60° C
Relative Humidity: 0-80%
How Supplied
The ENSEAL
G2 Tissue Sealer with integrated power cord is supplied sterile for single patient use.
®
Discard after use.
Warnings and Precautions
Caution: Federal (USA) law restricts this device to sale by or on the order of a physician.
Examine the shipping carton and instrument for signs of shipping damage. Note any shortages,
breakage, or apparent damage, retain the evidence, and notify Ethicon Endo-Surgery Customer Service
immediately and replace with a new instrument. Do not use a damaged product.
A thorough understanding of the principles and techniques involved in laser, electrosurgical, and
ultrasonic procedures is essential to avoid shock and burn hazards to both patient and medical
personnel and damage to the device or other medical instruments. Ensure that electrical insulation
or grounding is not compromised.
Laparoscopic surgical procedures with the ENSEAL
persons with adequate training and preparation. Consult the medical literature regarding techniques,
complications, and hazards prior to any endoscopic procedure.
Do not immerse the ENSEAL
Pools of conductive fluids (e.g. blood, saline) in direct contact with the instrument jaws may affect
device performance. Remove pooled fluid from around the instrument jaws before activating the
instrument.
Verify compatibility with generators. Use device only with the Generator G11 (GEN11).
The device is at high voltage whenever the generator is active. Activate this device only when the jaws
are under direct visual observation to avoid unintended tissue coagulation or necrosis.
Check if patient has a pacemaker or implanted cardioverter/defibrillator. Consult pacemaker
manufacturer for information about the effects of RF energy on these devices.
This instrument contains a surgically sharp edge. THIS DEVICE CUTS!
(Illustration 1)
6. Energy Activation Button
7. Closing Handle
8. Grip Housing
9. Power Cord
G2 Tissue Sealer in liquid.
®
3
G2 Tissue Sealer should be performed only by
®