EMC Information

The following EMC information and its associated tables replace EMC content comtained in Chapter 13 of the
Trilogy manual.
Your unit has been designed to meet EMC standards throughout its Service Life without additional maintenance. There is
always an opportunity to relocate your device within an environment that contains other devices with their own unknown
EMC behavior. If you believe your unit is affected by locating it closer to another device, simply separate the devices to
remove the condition.
Pressure and Flow Accuracy
This therapy device is designed to perform within the pressure and flow rate accuracies specified in the user manual. If you
suspect that the pressure and/or flow rate accuracy is affected by EMC interference remove power and relocate the device
to another area. If performance continues to be affected discontinue use and contact your home care provider.
SpO
and Pulse Rate Accuracy
2
This therapy device is designed to capture the SpO
described in the sensor manufacturer's instructions for use. When 4 hours of successful oximetry data have been achieved
the device indicates this to the user by displaying "Oximetry: Good Study". If you suspect that your unit is affected by EMC
interference remove power and relocate the device to another area. If performance continues to be affected discontinue
use and contact your home care provider.

Guidance and Manufacturer's Declaration - Electromagnetic Emissions

This device is intended for use in the electromagnetic environment specified below. The user of this device should
make sure it is used in such an environment.
e
t
missions
est
RF radiated emissions
CISPR 11
RF emissions
CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations/Flicker
emissions
IEC 61000-3-3
Emission of Radio Frequency
Energy
RTCA/DO-160G Section 21
2
c
omPliAnce
Group 1
Class B
Group 1
Class B
Class A
Complies
Category M
and Pulse Rate oximetry data within the accuracy specification
e
lectromAgnetic
The device uses RF energy only for its internal function. Therefore, its
RF emissions are very low and are not likely to cause any interference in
nearby electronic equipment.
The device is suitable for use in all establishments, including domestic
establishments and those directly connected to the public low-voltage
power supply network.
This device is suitable for use onboard commercial airplanes inside passenger
cabin.
e
- g
nVironment
uidAnce