Philips RESPIRONICS Trilogy Mode D'emploi page 13

Additif
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 28
A
P
lArm
riority
Low Battery
Escalates
from
Medium to
High
Replace
Low or
Detachable
High,
Battery
depending
on cause of
alarm
AC Power
Medium
Disconnected
Keypad Stuck
Low
A
V
udible
isuAl
i
ndicAtor
And
• • •
Medium Priority-Yellow
flashing button. "Low
(Medium -
Detachable Battery,"
when approx.
"Low External Battery,"
20 minutes
or "Low Internal Battery"
remains)
message appears in
yellow, On Status Panel,
box around battery is
yellow
High Priority-Red flashing
button. "Low Detachable
• • • • •
Battery", "Low External
(High - when
Battery," or "Low Internal
approx. 10
Battery" message appears
minutes
in red. On Status panel,
remains)
box around battery is red
• •
"Replace Detachable
Battery" message appears.
for Low
If battery is nearing end
• • • • •
of useful life, message
for High
appears with yellow
background and button is
solid yellow. If battery fails,
message appears with red
background and button
flashes red.
• • •
Yellow flashing
button; "AC Power
Disconnected" message,
and a box appears around
battery in use.
• •
Solid yellow button;
"Keypad Stuck" message.
(A
d
lArm
eVice
b
A
utton
ction
d
)
isPlAy
Operates
Operates
Switches
to
alternate
power
source
Operates
u
/c
ser
liniciAn
User: Press Alarm Indicator/Audio Pause
button to silence alarm. Or, press Left button
(Reset) to reset alarm. Switch to an alternate
battery. Or, switch to AC power source and
recharge low battery.
Clinician: Switch to alternate battery or AC
power while you recharge low battery. If low
battery is recharged and alarm continues,
replace battery.
User: Press Alarm Indicator/Audio Pause
button to silence alarm. Or, press Left button
(Reset) to reset alarm. Switch to an alternate
battery or AC power source while replacing
current detachable battery.
Clinician: Switch to an alternate battery or
AC power source while replacing the current
detachable battery.
User: Press Alarm Indicator/Audio Pause
button to silence alarm. Or, press Left (Reset)
button to reset alarm. If you were using AC
power, check power cord and reconnect it if
it has become disconnected. Make sure device
is not connected to an overloaded AC circuit.
If you were using an external, detachable, or
internal battery, recharge the battery.
Clinician: Check batteries and recharge if
necessary. Restore AC power if available.
User: Press Alarm Indicator/Audio Pause
button to silence alarm. Or, press Left (Reset)
button to reset alarm. Check the keys to
determine if they are lodged in the case. If
alarm continues, place patient on alternate
source of ventilation and contact your health
care professional.
Clinician: Check keys to determine if they
are lodged in the case. If alarm continues,
place patient on alternate source of
ventilation and have device serviced.
A
ction