Patient Alarms
A
P
lArm
riority
Check Circuit
High
Low Circuit
High
Leak
High
High
Expiratory
Pressure
Low
High
Expiratory
Pressure
A
V
udible
isuAl
i
ndicAtor
And
• • • • •
Red flashing button;
"Check Circuit" message
• • • • •
Red flashing button;
"Low Circuit Leak"
message
• • • • •
Red flashing button;
"High Expiratory
Pressure" message
• • • • •
Red flashing button;
"Low Expiratory
Pressure" message
(A
d
lArm
eVice
b
A
utton
ction
d
)
isPlAy
Operates
Operates
Operates
Operates
u
/c
ser
liniciAn
User: Press Alarm Indicator/Audio Pause
button to silence alarm. Or, press Left
(Reset) button to reset alarm. Make certain
the circuit type matches what is selected
in the Setup Menu. Make sure there is no
kinked or pinched tubing. Make sure tubing
is properly attached. Make sure all tubing
and exhalation devices are clean and free of
condensation (refer to the Troubleshooting
section - tubing condensation). If using Active
PAP or Active Flow Exhalation Device, make
sure diaphragm is inserted correctly and is
not wrinkled or kinked. If alarm continues,
contact your health care professional.
Clinician: Verify patient status. Verify that
breathing circuit setup is correct. Correct any
errors. If alarm continues, have device serviced.
User: Press Alarm Indicator/Audio Pause
button to silence alarm. Or, press Left (Reset)
button to reset alarm. Check for occlusions
in exhalation devices. Make sure exhalation
device is clean and functioning properly.
Refer to the Troubleshooting section if Low
Circuit Leak alarm is sounding during nebulizer
treatments. If alarm continues, contact your
health care professional.
Clinician: Remove obstruction in leak device.
If alarm continues, have the device serviced.
User: Press Alarm Indicator/Audio Pause
button to silence alarm. Or, press Left (Reset)
button to reset alarm. Make sure tubing is not
kinked or pinched. Check patient's breath rate.
If alarm continues, contact your health care
professional.
Clinician: Check patient's tubing to make sure
it is not kinked or pinched. If alarm continues,
have device serviced.
User: Press Alarm Indicator/Audio Pause
button to silence alarm. Or, press Left (Reset)
button to reset alarm. Make sure tubing is not
kinked or pinched. If alarm continues, contact
your health care professional.
Clinician: Check patient's tubing to make sure
it is not kinked or pinched. If alarm continues,
have device serviced.
A
ction