WARNING
Leave the area immediately. You are in mortal danger! A main
alarm is self-latching and cannot be acknowledged or can-
celled.
After leaving the area, if the concentration is less than the alarm
threshold A2:
Press the [OK] key, the alarm messages are switched off.
STEL / TWA exposure alarm
The alarm is indicated by an intermittent
alarm message:
Display » A2 « and »
« (STEL) or »
ured value alternating:
CAUTION
Leave the area immediately. After this alarm, the deployment
of personnel is subject to the relevant national regulations.
– The STEL and TWA alarm cannot be acknowledged or can-
celled.
Switch off the device. The values for the exposure evaluation
are deleted after the device is switched on again.
Battery pre-alarm
The alarm is indicated by an intermittent
alarm message:
Flashing special symbol »
« on the
right side of the display:
Acknowledging the pre-alarm:
Press the [OK] key, only the audible alarm and the vibration
alarm are switched off.
– The battery still lasts approx. 20 minutes after the first battery
pre-alarm.
Battery main alarm
The alarm is indicated by an intermittent
alarm message:
Flashing special symbol »
« on the
right side of the display:
The battery main alarm cannot be acknowledged or
cancelled:
– The device is automatically switched off again after
10 seconds.
– Before the device is switched off, the visual, audible and
vibration alarms are activated for a short time.
Device alarm
The alarm is indicated by an intermittent
alarm message:
Special symbol »
right side of the display:
– The device is not ready for operation.
Commission maintenance personnel or the Dräger Safety
Service Centre to eliminate the error.
Calling the Info Mode
In measuring mode, press the [OK] key for approx. 3 sec-
onds.
« (TWA) and meas-
– If any warning or fault messages exist, the corresponding
note or error codes are displayed (see the Technical Hand-
book).
Press the [OK] key successively for the next display. The
peak values and the exposition values TWA and STEV will be
displayed.
– If no key is pressed for 10 seconds, the device returns auto-
matically to measuring mode.
Info Off Mode
With the device switched off, press the [ + ] key. Gas name,
measuring unit and measuring range limit value will be dis-
played for all channels.
Pressing the [ + ] key again exits the Info Off Mode (or via time-
out).

Calling the Quick Menu

In measuring mode, press the [ + ] key three times.
– If functions in the quick menu are activated using the PC soft-
ware "Dräger CC-Vision", you can select these functions us-
ing the [ + ] key. If no functions have been activated in the
quick menu, the device remains in measuring mode.
Possible functions:
Press the [OK] key to call the selected function.
Press the [ + ] key to cancel the active function and to switch
to measuring mode.
– If no key is pressed for 60 seconds, the device returns auto-
matically to measuring mode.
« displayed on the
1. Bump test mode
2. Fresh air calibr.
3. Display and deletion of the
peak values
Replacing the Batteries /
Rechargeable Batteries
WARNING
Do not replace the batteries / rechargeable batteries in haz-
ardous areas. Danger of explosion!
Batteries / rechargeable batteries are part of the Ex approval.
Only the following types may be used:
– Alkaline batteries – T4 – (not rechargeable!)
Energizer No. E91, Energizer No. EN91 (Industrial),
Varta Type 4106 (power one) or
Varta Type 4006 (industrial)
– NiMHy rechargeable batteries – T3 – (rechargeable)
o
GP 180AAHC (1800) max. 40
C ambient temperature.
Switch off the device:
Press and hold the [OK] key and the [ + ] key simultaneously.
Loosen the screw on the power pack and remove the power
pack.
For the battery holder (order no. 83 18 703):
Replace alkaline batteries or NiMHy rechargeable batteries.
Ensure correct polarity.
For the NiMH power pack T4 (order no. 83 18 704):
Completely replace the power pack.
Insert the power pack into the device and tighten the screw,
the device switches on automatically.
WARNING
Do not throw used batteries into fire or try to open them by
force. Danger of explosion!
Dispose of the batteries in accordance with local regulations.
Charge the device with NiMH power
pack T4 (83 18 704)
WARNING
Do not charge underground or in explosion-hazard areas!
Danger of explosion!
The chargers are not designed in accordance with the regula-
tions for firedamp and explosion protection.
Even if the device is not used, we recommend that you store the
device in the charger!
Insert the switched off device into the charger.
– Display LED on the charger:
Charge
Fault
Full
9