Factory default settings

12.5 Sensor test

This function is only active if the sensor test dongle is
installed. If the sensor does not pass the self-test, a warning
or fault message is created.
Setting periodic sensor test
This function periodically initiates the sensor self-test. The
instrument routinely checks the sensor for proper function.
1. Select Settings > Sensor > Set sensor test and confirm.
2. Select Enable or Disable and confirm.
Starting sensor test manually
This function starts a singular sensor test.
1. Select Settings > Sensor > Sensor test and confirm.
2. Select Start sensor test and confirm.
3. Select Confirm
✓ The test result is displayed.

12.6 Gas settings

Gas settings are not used in Polytron 8900 UGLD.
13
Factory default settings
13.1 Fixed settings for relays, LED and
alarm
Setting
Fault relay
Energized / Normally Open
Yellow Fault LED
Lit if a fault message is available
Red Alarm LED
Single blink if A1 condition is present.
Double blink if A2 condition is
present. If an alarm is configured as
acknowledgeable and the alarm is
acknowledged, single/double blink
changes into continuously lit.
Alarm Hierarchy
A2 overrides A1 on the LED. However,
the A1 and A2 relays operate inde-
pendently. I.e. if A1 is acknowledgeable
and A2 is not, and the gas concentra-
tion is such that it triggers A1 and A2:
acknowledging will cause the A1 relay
to release. However, the red LED will
still double blink as long as the A2 con-
dition continues to exist.
13.2 Settings which can be changed via
the menu
Menu
Default setting
Relay active on
Active on alarm
alarm / No alarm
22
Menu
A1 alarm: direction Rising
A2 alarm: direction Rising
A1 latch mode
A2 latch mode
A1 acknowledg-
ment mode
A2 acknowledg-
ment mode
Normal operation
A1 relay
Normal operation
A2 relay
Calibration pass-
word
Settings password _ _ _ 2
LCD setting
SIL status
Language
Function key
Interface
Fault current
Warning
Warning current
Warning cycle time
T1 (measurement
current)
Warning cycle time
T2 (warning cur-
rent)
Maintenance sig-
nal
Maintenance cur-
rent
Analog offset
Analog span
HART address

13.3 Sensor specific values

Range
Menu
On / Off
A1 Alarm
A2 Alarm
Default setting
Non-latching
Latching
Acknowledgeable
Not acknowledge-
able
Energized
_ _ _ 1
On
Off
EN
Faults
Default setting
1.2 mA
Off
3.0 mA
9 s
1 s
static
3.4 mA
0 mA
0 mA
0
Default setting
99 %
100 %
Instructions for use
|
Dräger Polytron
Range
Falling / Rising
Latching / Non-
latching
Acknowledgeable /
Not acknowledge-
able / Pre-acknowl-
edgeable
Energized / Not
energized
On / Off
On / Off
DE / EN / FR / ES /
RU / ZH
Graph, fault, warn-
ing, vitality, bump
test
Range
0 to 3.5 mA
On / Off
0 to 3.5 mA
5 to 60 s
1 to (T1-1) s
static / dynamic
0 to 3.5 mA
-0.5 to 0.5 mA
-0.5 to 0.5 mA
0 to 15
Range
100 % (110 dB)
®
8900 UGLD