Implementing a reset
1. Hold down the following three buttons simultaneously
for about three seconds on the hand control: Red
STOP button A, Forward button B and Up button C.
An acoustic signal will be heard while the three buttons
are held down.
2. When „SERVICE" indicator lamp D goes out, release
the buttons. The device is ready to use again.
3. Then move the drives for adjusting the height to the
lowest position until you hear the drives switch off.
4. Move the tilt drive to the rear end position until you
hear the drive switch off. (The red LED flashes and an
acoustic signal is heard).
Replace the safety clips with new ones if the
actuators are replaced.
Make sure that all cables are re-installed the same
way as supplied if components have to be changed.