5.4.1 Replacing the flexible mounting pin

Phillips screwdriver
1. Remove the flexible mounting pin A by unscrewing the
screw B on the underside of the foot plate C.
2. Remove the heel strap D.
To install the flexible mounting pin reverse steps 2
– 1.
B = 4 Nm

5.4.2 Removing/Installing the swivel joint

In case the legrest tubes must be cut to length, the swivel joints must be removed.
Allen key (6 mm)
1. Remove bolt A and nut B.
2. Remove the swivel joint C from the legrest tube D.
To install the swivel joint reverse steps 2 – 1.
A = 4 – 5 Nm

5.5 Replacing the castors

Allen key (10 mm) / Thread locking adhesive (e.g. Loctite 2701)
1. Remove the castor from the side part by unscrewing
bolt A from the threaded hole B.
To install the castor reverse step 1.
A = 35 Nm, apply thread locker e.g. Loctite 2701