Information
The program counter and the operating hours can be viewed
in the information area. The values can be reset by Dräger
Service if necessary.
NOTICE
A wash procedure is counted as soon as it is started. The wash
procedure does not have to be carried out completely in order
to count.
Program counter The number of times a wash procedure
has been started is displayed.
Operating hours
How long different components were
operated is displayed in hours.
4.3.7
System settings
The following settings can be made in the system settings.
Plant mode
The plant mode can be selected by the
Admin user between Manual and Auto. In
Manual plant mode, all valves can be
operated manually. If a wash procedure is
started when the MFC 7000 is in Manual
mode, the mode is automatically switched
to Auto.
Manual plant mode is used to test certain
components for maintenance work or for
troubleshooting. If a wash procedure is
active and the plant mode is switched
from Auto to Manual, the active wash
procedure is aborted.
44
General
If no wash procedure is active, the current
user can be logged off with the
The
logging in again.
The
language for the software.
Calibrate touch-
The
screen
touchscreen.
Set time
Tap on the blue "New time" box and set
the new time and date. The settings must
be made in MM/DD/YYYY h:mm:ss AM/
PM format; otherwise they will not be
saved. Confirm the settings with the
button.
4.3.8
Alarm message
The alarms view can be accessed by pressing the F6 key
below the touchscreen. The alarms view lists all pending
alarms.
Time
Time at which the error occurred.
Date
Date at which the error occurred.
Status
Status of the error
K - Coming, new error present.
RP - Going, rectified error.
Q - Acknowledged, acknowledged error
present.
Text
Brief description of the error
As soon as an alarm has been marked, the
explanation of the alarm with the fault remedy
can be called up by pressing the button.
button then appears, to allow
button is used to change the
button is used to recalibrate the
button.