Invacare® Robin™
Read Feeding Voltage
Take voltage feeding the trolley
mover motor whilst pressing
the LEFT or RIGHT button
on the hand control. (Robin
Mover™ only)
Expected result: 23 to 29 V
Take voltage feeding the
solenoid whilst pressing the
LEFT or RIGHT button on the
hand control. (Robin Mover™
only)
Expected result: 23 to 29 V
28
1586747-A