Replacing The Hand Control; Removing The Hoist Cover; Replacing Strap And Hook Assemblies - Invacare Robin Manuel De Maintenance

Masquer les pouces Voir aussi pour Robin:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 65
Invacare® Robin™

3.3 Replacing the hand control

Slide and hold locking lever A away from the aperture C.
1.
Pull the plug B out of the aperture.
2.
3.
Insert plug of the new hand control into aperture. Make sure,
the plug is in the correct orientation with the slot upwards (see
image).
4.
Release lever A, so that the gate closes and hold the plug in
the aperture.

3.4 Removing the hoist cover

Before removing the hoist cover, pull the red emergency strap to turn
off the hoist and remove the hoist from the track system. Refer to:
3.2 Dismounting the ceiling hoist, page 9
Tools: Screw driver, slotted
12
Remove protection caps A. Use fingers or, if required, use a
1.
screw driver carefully.
Untighten screws B, using a screw driver.
2.
Remove the cover C from the hoist.
3.
To close the hoist cover, tighten the screws and attach the protection
caps again.

3.5 Replacing Strap and Hook Assemblies

Tools: 10 mm socket key; 5 mm hex/Allen key
1.
With the ceiling hoist in the track, lower the straps to their
lowest point using the hand control.
2.
Pull the red emergency strap to perform an emergency stop. Do
not perform emergency lowering.
3.
Remove hoist from track. Refer to 3.2 Dismounting the ceiling
hoist, page 9 .
4.
Remove the cover. Refer to 3.4 Removing the hoist cover, page
12.
1586747-A

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

Robin mover

Table des Matières