Invacare® Robin™

6.3 Electrical System

Robin™
Maximum current
input
1
Protection class
(ceiling hoist: IP24; hand control: IP44)
Insulation class
Applied part complying with the specified
requirements for protection against electrical
shock according to IEC60601-1.
Sound level
Number of lifts per
charge
Intermittent
(periodic motor
operation)
Battery
Charging time
Lifting speed with
2.1 cm/s
200 kg load
Traverse speed in
the tracks
30
Robin Mover™
0.9 A
Entire device: IPx4
Class II equipment
Type B applied part
50 – 55 dB (A)
65 lifts of 0.5 m with 80 kg
10%, max. 2 min. / 18 min.
NiMH Battery 24V / 2.0 Ah
Up to 3 hours at 20°C
2.1 cm/s
n/a
14 cm/s
Charger
Voltage output
charger
Voltage supply
charger
1
Protection class
IP41 or IP67 (depending on version)
charger
1
See product label and label on each electric device for correct
protection class. The lowest IP classification decides the overall
classification of the device.
IPx4: Protected against water splashed from any direction.
IP24: Protected against objects larger than 12.5 mm and
protected against water splashed from any direction.
IP44: Protected against objects larger than 1 mm and protected
against water splashed from any direction.
IP41: Protected against objects larger than 1 mm and protected
against vertically dripping water.
IP67: Dust-tight and protected against temporary immersion
into water.

6.4 Environmental conditions

Storage and
transportation
Temperature
-30 °C to +50 °C
Relative humidity
10 % to 70 %
Atmospheric
pressure
24 V DC, max. 250 VA
230 V AC, 50 Hz
Operation
+5 °C to +40 °C
20 % to 90 % at 30
°C, not condensing
700 hPa to 1060 hPa
1586747-A