5.3
Setting up and connecting the centrifuge
Setting up the centrifuge
AB4701en
5
1
4
2
4
5
Fig. 21: Transport lock
1
Screws
2
Motor cover
3
*Bellows (only for centrifuges with cooling)
4
Transport lock
5
Transport lock screws
6
Spacer sleeve
1.
Open the lid.
2.
Unscrew 4 screws (
3.
Remove the motor cover (
4.
Unscrew 3 transport lock screws (
5.
Remove the transport lock screws (
4 ) and store them safely.
port locks (
6.
Turn over the motor cover (
cable.
7.
For ROTINA 420 R:
3 ) over the edge of the centrifuging chamber.
Push the bellows (
8.
Screw in 4 screws (
WARNING
Risk of injury
Due to insufficient clearance from the centrifuge.
−
As per EN / IEC 61010-2-020, no persons, hazardous
materials or objects may be present within a safety zone
of 300 mm around the centrifuge during a centrifugation
run.
−
Maintain a distance of 300 mm from the ventilation slots
and ventilation openings of the centrifuge.
−
The centrifuge's ventilation openings must never be
blocked.
Rev.: 19 / 02.2025
1
3
4
1 ).
2 ).
5 ).
5 ), spacer sleeves ( 6 ) and trans-
2 ) and insert it. Note the recess for the
1 ).
Commissioning
1
2