Device overview
Mains frequency
power consumption
Power consumption
Refrigerant
max. capacity
max. permissible density
max. speed (RPM)
max. acceleration (RCF)
max. kinetic energy
Obligation to perform
checks
(DGUV Rules 100-500)
(valid only in Germany)
Ambient conditions (EN / IEC 61010-1):
Installation site
Altitude
Ambient temperature
Humidity
IP ratings
Overvoltage category
(IEC 60364-4-443)
Pollution level
Device protection class
EMC:
Emitted EM interference,
EM interference immunity
Noise level
(rotor-dependent)
Dimensions:
Width
Depth
12 / 64
50 Hz
60 Hz
1600 VA
1300 VA
7.5 A
R452A
4 x 600 ml
1.2 kg/dm³
15000
24400
24000 Nm
yes
indoors only
up to 2000 m above sea level
5 °C to 35 °C
maximum relative humidity 80% for temperatures up to 31 °C,
decreasing linearly to 50% relative humidity at 40 °C.
IP 20
II
2
I
not suitable for use in potentially explosive atmospheres.
EN / IEC 61326-1
Class B
£58 dB(A)
£64 dB(A)
713 mm
654 mm
Rev.: 19 / 02.2025
60 Hz
50 Hz
1850 VA
16.0 A
FCC Class B
£58 dB(A)
AB4701en